Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
117
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Air/water channel cleaning adapter (MAJ-629)
During precleaning, the air/water channel cleaning adapter is connected to the
air/water cylinder. When the adapter is depressed, water is fed through the
air/water channel. Air is continuously fed when the adapter is not depressed (see
Figure 7.10).
Figure 7.10
Use the air/water channel cleaning adapter only during
precleaning.
Injection tube (MH-946)
The injection tube is used to inject detergent solution, disinfectant solution, water
and alcohol into the air/water channel, instrument channel and suction channel
and to flush air through the channels to expel fluids (see Figure 7.11).
Figure 7.11
Suction channel tube
Air pipe port
Air/water channel tube
Connector plug
Notice card
Suction channel port
Air/water channel port
Suction channel port
(including filter mesh)