Important Information — Please Read Before Use
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Do not attempt to bend the endoscope’s insertion section
with excessive force. Otherwise, the insertion section may be
damaged.
Do not touch the electrical contacts inside the videoscope
cable connector. CCD damage may result.
Do not apply shock to the distal end of the insertion section,
particularly the ultrasound transducer and the objective lens
surface at the distal end. Visual abnormalities may result.
Do not hold the ultrasonic transducer when holding the
insertion tube. The ultrasonic transducer damage can result
and/or the ultrasonic image will be abnormal.
Do not twist or bend the bending section with your hands.
Equipment damage may result.
Do not squeeze the bending section forcefully. The covering
of the bending section may stretch or break and cause water
leaks.
Turn the video system center ON only when the videoscope
cable is connected to both the video system center and the
videoscope cable connector on the endoscope. In particular,
confirm that the video system center is OFF before
connecting or detaching the videoscope cable from the
electrical connector on the endoscope. Failure to do so can
result in equipment damage, including destruction of the
CCD.
The endoscope’s remote switches cannot be removed from
the control section. Pressing, pulling, or twisting them with
excessive force can break the switches and/or cause water
leaks.
If remote switch 1 does not return to the OFF position after
being pressed strongly from the side, gently pull the switch
upwards to return it to the OFF position.
Do not hit or bend the electrical contacts on the endoscope
connector. The connection to the light source may be
impaired and faulty contact can result.
Do not touch the electrical contacts in the ultrasonic cable
connector. Equipment damage can result.
Do not pull, twist or tightly coil the ultrasonic cable. Noise can
develop in the ultrasound image.