Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
8. Using the 30 cm
3
(30 ml) syringe, flush the air/water, balloon, and suction
channels with 150 cm
3
(150 ml) of air, respectively, via the injection tube
(see Figures 9.19 and 9.22).
9. Use the 5 cm
3
(5 ml) syringe, to flush the elevator wire channel with 10 cm
3
(10 ml) of air via the washing tube (see Figure 9.21).
10. Remove the clean, lint-free cloth from the distal end and the control section
of the endoscope.
11. Detach the channel plug, the cleaning adapter for instrument channel port
and injection tube from the endoscope.
12. Detach the washing tube from the endoscope.
Figure 9.22
Suction cylinder
Channel plug
Instrument channel
Balloon channel
Suction channel
Universal cord
Suction connector
Suction port
Syringe
Suction channel port
Air/water
channel port
Injection tube
Suction
channel tube