Chapter 13 Troubleshooting
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
EndoTherapy accessories
Others
Irregularity
description
Possible cause Solution
An EndoTherapy
accessory does not
pass through the
instrument channel
smoothly.
An incompatible
EndoTherapy accessory is
being used.
Refer to the “System chart” in the
Appendix and select a compatible
EndoTherapy accessory. Confirm
that the color code on the
EndoTherapy accessory matches
that on the endoscope. Although
the color code of the GF-UCT180 is
orange, do not use EndoTherapy
accessories designed for a
Ø 4.2 mm or Ø 6.0 mm channel,
use only EndoTherapy accessories
designed for a o 3.7 mm channel.
The elevator
control lever does
not operate
smoothly.
The elevator wire or the
elevator wire channel is
dirty.
Clean and disinfect or sterilize the
elevator wire channel as described
in Chapter 7, “Cleaning,
Disinfection, and Sterilization
Procedures” that has the model
number of your endoscope on the
cover.
Irregularity
description
Possible cause Solution
The remote switch does
not work.
The wrong remote
switch is operated.
Operate the correct remote
switch.
The remote switch
function has been set
incorrectly.
Set the remote switch function
correctly as described in the video
system center’s instruction
manual.