Appendix C: EMC Information
233
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
• Electromagnetic interference may occur in the vicinity of
high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Immunity
test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic environment —
Guidance
Conducted RF
IEC 61000-4-6
3V
(150 kHz –
80 MHz)
3V (V
1
)
(150 kHz –
80 MHz)
Same as left Recommended separation distance
Where “P” is the maximum output power
rating of the transmitter in watts [W]
according to the transmitter manufacturer
and “d” is the recommended separation
distance in meters [m].
6V
(ISM band of
150 kHz –
80 MHz)
–
Same as left
ISM (industry, science, and medical care) band of 6.765 MHz – 6.795 MHz, 13.553 MHz – 13.567
MHz, 26.957 MHz – 27.283 MHz, and 40.66 MHz – 40.70 MHz between 0.15 MHz and 80 MHz
Radiated RF
IEC 61000-4-3
3V/m (80 MHz –
2.7 GHz)
3V/m (E
1
)
(80 MHz –
2.5 GHz)
Same as left Recommended separation distance
80 MHz – 800 MHz
800 MHz – 2.5 GHz
Where “P” is the maximum output power
rating of the transmitter in watts [W]
according to the transmitter manufacturer
and “d” is the recommended separation
distance in meters [m].
Proximity
magnetic field
from RF
communication
equipment
IEC 61000-4-3
Refer to the table
of the next page.
–
Same as left