Chapter 2 Instrument Nomenclature and Specifications
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Connecting the fitting part to the suction connector may
impair the connection of the suction tube to the suction
connector. It may also cause the suction tube to become
detached from the endoscope and allow patient debris to
spray.
When attaching the water-resistant cap to the connector of
the endoscope, do not pinch the chain for water-resistant cap
between the connector of the endoscope and the
water-resistant cap. Otherwise, equipment damage may
result.
The chain for water-resistant cap and water-resistant cap
cannot be steam sterilized (autoclaved); doing so can
damage them severely.
Figure 2.1
To ensure that you do not forget to attach the water-resistant
cap, it is recommended that you connect the chain for
water-resistant cap to the endoscope’s S-cord connector
mount.
1. Confirm that the chain for water-resistant cap is free from cracks, flaws,
wear, deformation, or other damages (see Figure 2.1).
2. Align the notch on the connecting plate with the pin on the venting connector
of the water-resistant cap (MH-553, see Figure 2.2).
3. Place the connecting plate over the venting connector (see Figure 2.2).
4. Confirm that the connecting plate is securely attached to the foot of the
venting connector and can be smoothly rotated.
Notch
Chain part
Fitting part
Connecting plate
Hole