Chapter 4 Operation
61
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Insertion of the endoscope
Keep the elevator control lever moved all the way in the
opposite direction of the “ Udirection while inserting or
withdrawing the endoscope into or from the patient. If the
elevator control lever is moved all the way in the “ U”
direction and the forceps elevator is raised while inserting or
withdrawing the endoscope into or from the patient, this may
cause patient injury.
To prevent the patient from accidentally biting the insertion
section during an examination, it is strongly recommended
that a mouthpiece be placed in the patient’s mouth before
inserting the endoscope.
To prevent the patient from accidentally loosing dental
prosthesis, make sure that the patient removes it before the
examination.
Do not apply olive oil or products containing petroleum-based
lubricants (e.g., Vaseline
®
) to the endoscope. These
products may cause stretching and deterioration of the
bending section’s covering.
Do not allow the insertion section to be bent within a distance
of 10 cm or less from the junction of the boot. Insertion
section damage can occur (see Figure 4.2).
Figure 4.2
> 10 cm