2 Rotate the check tool so that you can get an image similar to the one shown in the
picture below.
Side-view optical adapter
Forward-view optical adapter
3 Press the [MEAS/OK] joystick or push the [ZOOM] lever downward while the live
screen is displayed.
The spot ranging screen is displayed.
4 While the distal end of the insertion tube is being inserted into the stereo optical
adapter hole of the check tool until it touches the end, confirm that the distance is 15
mm or less in [Object distance] and press the [MEAS/OK] joystick.
The measurement method selection screen is displayed.
5 Select [DISTANCE] and press the [MEAS/OK] joystick.
The measurement screen is displayed.
6 Operate the [MEAS/OK] joystick to specify the measuring points (2 points)
7 Long-press the [FRZ/REC] button.
The measured image is recorded.