3
Registering/re-registering a stereo optical adapter
Before performing stereo measurement using a stereo optical adapter, register the stereo optical
adapter to this unit.
Similarly, if the stereo optical adapter and insertion tube combination has changed due to
reasons such as purchase of a new optical adapter or repair of the insertion tube, register the
stereo optical adapter again.
Also, if the measured value is found to be incorrect upon checking the result, go through the
steps of a new registration again to re-register the stereo optical adapter.
This registration process is to create information on the combination of the stereo optical adapter
and insertion tube that you are using.
Information that is created will be stored in the built-in memory of this unit.
Registering a new stereo optical adapter
1 Attach the stereo optical adapter to the distal end of the insertion tube.
See “Attaching the stereo optical adapter” (page 76) for details about operations.
2 In “Menu Operations”, select “SELECT OPTICAL ADAPTER” followed by “NEW STEREO
ADAPTER” and press the [MEAS/OK] joystick.
3 Select whether to use a forward view or a side view type.
77