Why does a message appear while using the high beam during inspection on Olympus Analytical Instruments?
S
Samantha PowellSep 9, 2025
If a message appears while using the high beam during inspection with your Olympus Analytical Instruments (LX only), it could be due to:
* The distal end of the insertion tube being too hot. Cool the distal end of the insertion tube to use the high beam.
* The distal end of the insertion tube is too hot, triggering the self- check function.
* The inside of the main unit is too hot, triggering the self- check function. Cool the inside of the main unit to use the high beam.
M
michael20Sep 8, 2025
Why is the operation load for articulation getting heavier on Olympus IPLEX LX Analytical Instruments?
C
Cory HillSep 8, 2025
If the articulation is getting heavier on your Olympus Analytical Instruments, it may be due to the insertion tube being overloaded, triggering the self-check function. To solve this, extend the insertion tube as much as possible to reduce the loop amount, and turn the power back on without operating the [ANGLE] joystick.
J
Jeffrey HoltSep 7, 2025
What to do if the image cannot be adjusted on Olympus Analytical Instruments?
D
Daniel PerezSep 7, 2025
If the image cannot be adjusted on your Olympus Analytical Instruments, the self-check function might be triggering during use and displaying a message. To resolve this, stop the inspection and turn the instrument off, then on again.
J
Joe IrwinAug 22, 2025
What to do if the image cannot be recorded or retrieved on Olympus IPLEX LX Analytical Instruments?
D
David KingAug 22, 2025
If you are unable to record or retrieve images with your Olympus Analytical Instruments, there might be several reasons:
* Ensure a USB Flash Drive is properly inserted and restart the device.
* If the USB Flash Drive is full, delete unnecessary files or use a new one.
* Only images recorded with the IPLEX LX/LT can be retrieved.
* Format the USB Flash Drive with the IPLEX LX/LT if the format is not recognized.
* If the self-check function is activated, stop the current process, turn off the instrument, and then turn it back on.
P
Pedro LeonSep 5, 2025
Why is the illumination dim on my Olympus Analytical Instruments?
C
chapmankellySep 5, 2025
If the illumination appears dim on your Olympus Analytical Instruments, consider these factors:
* The optical adapter tip is dirty. Clean it with a clean piece of gauze or cotton swab.
* Extended use under high ambient temperatures causes decreased light emissions. In this case, replace the optical adapter.
D
Daniel JamesSep 3, 2025
Why the optical adapter cannot be attached to the insertion tube of Olympus IPLEX LX?
J
Joan WileySep 4, 2025
If the optical adapter cannot be attached to the insertion tube on your Olympus Analytical Instruments, check for these issues:
* A foreign object is stuck to the screws. Wipe them with a clean piece of gauze or cotton swab.
* The optical adapter in use is not a designated optical adapter for this system. Use a designated optical adapter.
* You are not using the proper installation procedure. Redo it correctly.
D
Douglas AyersSep 2, 2025
What to do if the Olympus IPLEX LX Analytical Instruments system cannot be turned off?
H
hdavisSep 2, 2025
If your Olympus Analytical Instruments system cannot be turned off, it suggests that the main unit is damaged. To turn off the power, remove the AC adapter or the battery from the main unit.
M
melanie15Sep 1, 2025
What does it mean when self-check triggers during use and displays a message on Olympus IPLEX LX?
M
Michele HooverSep 1, 2025
If the self-check function triggers during use and displays a message on your Olympus Analytical Instruments, it could indicate an abnormality was detected in the tip electrode of the insertion tube. To resolve this, immediately withdraw the insertion tube from the object of observation and turn the power off. Clean the insertion tube tip and optical adapter electrodes, and then reinstall the optical adapter.
N
Nicholas PerezAug 31, 2025
Why is the image brightness not optimum on my Olympus IPLEX LX Analytical Instruments?
J
Jon HoodAug 31, 2025
If the image brightness is not optimum on your Olympus Analytical Instruments, it could be due to several factors:
* The automatic brightness control is set erroneously. Perform the [BRT] lever operation to configure the proper setting.
* The object lens on the insertion tube distal end or the optical adapter is dirty, or the illumination on the optical adapter distal end is dirty. Clean them with a clean piece of gauze or cotton swab.
* The external monitor is adjusted incorrectly. Correctly adjust the external monitor.
* The optical adapter is old. Consider replacing it.
K
Katherine FullerAug 29, 2025
What does it mean when a message is displayed when using the battery on Olympus IPLEX LX Analytical Instruments?
T
Tina EspinozaAug 29, 2025
If a message appears while using the battery with your Olympus Analytical Instruments, it means the battery voltage has dropped. You should stop operations such as recording images, copying, deleting, or formatting a USB flash drive, and replace the battery immediately. Alternatively, attach the AC adapter.