• Do not install this equipment in any place where any of the following are present.
High oxygen concentration
Oxidizing substance such as Nitrous Oxide (N
2
O)
Flammable anesthetic gas
This equipment is not explosion-proof and may explode or cause a fire under these
conditions.
• Always use the power cord provided with this equipment. Otherwise, equipment
failure or power cord burnout may result. Also, remember that the provided power
cord is for use only with this equipment and should not be used with any other
equipment.
• Do not press any of the switches on the control panel of this equipment with a
pointed or hard object. Otherwise, the switch may be damaged.
• Be sure to turn off the water faucet and the power switch of the equipment at the
end of the day to avoid potential water leaks.
• To avoid malfunctions, do not use this equipment in a dusty environment.
• To avoid electromagnetic interference from other equipment, do not install any
other electrical devices in close proximity to this equipment (aside from ancillary
devices used with this equipment).
• This equipment enables radio communication by RFID and emits RF (radio
frequency:13.56MHz) energy to perform the said intended Functions. It may cause
electromagnetic interference in nearby electronic equipment, and is labeled with
the symbol below. If electromagnetic interference occurs, mitigation measures may
be necessary, such as moving the electronic equipment away, reorienting or
relocating this instrument, or shielding the location. An electromagnetic
interference with other devices may shorten the communications distance of the
designated ID tag and cause signals to become unreadable. Try to take mitigation
measures such as keeping the affecting device away from this equipment.