40
3.4 Inspection of the mesh filters
OER-Elite INSTALLATION MANUAL
Ch.3
The mesh filters are used to prevent the penetration of debris into the equipment and endoscope
channels.
CAUTION
If the mesh filters have been removed, be sure to put them back in their original
positions before using the equipment. If you forget to attach the mesh filters, the
pump may malfunction and/or foreign objects may clog the endoscope channels
including the nozzle.
After removing any of the mesh filters, be sure to reinstall them in their original
locations before using the equipment. Otherwise, debris may clog the endoscope
channels and nozzles or the pumps may fail.
If a mesh filter is dropped or impacted in any way, ensure that it is not damaged.
Damaged mesh filters may not perform adequately.
Two-type mesh filters are installed on the outer and inner sides of the circulation
port. Be sure to remove, inspect, and clean both of them.
3.4 Inspection of the mesh filters
1 Step on the foot pedal to open the lid.
2 Check that the drain port mesh filter and two circulation port mesh filters are attached
inside the reprocessing basin.
Figure 3.3
Drain port
mesh filter
Circulation port
mesh filters