60
4.5 Connection of the drain hose
OER-Elite INSTALLATION MANUAL
Ch.4
• Do not bend the drain hose or position it in such a way that it kinks and obstructs
access to the floor drain hole. Otherwise, discharge will be incomplete and the
endoscope may not be properly cleaned and disinfected.
• Connect the connection ring of the drain hose to the drain hose connection port on
the rear (bottom right) of the equipment by tightening securely. An insecure
connection may lead to a risk of a solution leak.
• Fix the other end of the drain hose firmly inside the facility’s floor drain hole.
Otherwise, the drain hose may come out of the hole due to the discharge pressure
resulting in fluid overflow.
• Install the drain hose so that its maximum height is 60 cm (23 in) or less.
Otherwise, discharge failure may cause the equipment to malfunction.
If the drain hose is too long, cut it to an appropriate length.
4.5 Connection of the drain hose
1 Align the connection ring of the drain hose with
the drain hose connector on the rear (bottom
right) of the equipment and turn the ring
clockwise to tighten it.
Figure 4.7
2 Insert the other end of the drain hose into the drain hole on the floor and secure it so
that it does not come out of the drain hole during discharge.