4.4 Connection of the water supply hose
59
OER-Elite INSTALLATION MANUAL
Ch.4
4 After confirming that the lock lever is engaged,
pull the hose lightly to make sure the
connection is tight.
Figure 4.5
CAUTION
Connect the connection ring on the equipment end of the water supply hose to the
water supply hose connection port by tightening securely. An insecure connection
may lead to a risk of a water leak.
5 Align the connection ring on the equipment end
of the water supply hose with the water supply
hose connector on the rear right of the
reprocessing basin so the hose does not hinder
the opening and shutting of the lid, and turn the
connection ring clockwise to tighten.
Figure 4.6
6 Open the water faucet slowly while checking to make sure that water does not leak
from the hose connection area. If water leaks, close the water faucet and reconnect
the water supply hose.
Connection ring on the
equipment end