Equipment compatibility
Use this equipment in combination with ancillary equipment listed in “System chart” in
“Instructions-Operation Manual”. Using incompatible equipment may interfere with the proper
operation of this equipment and could lead to personal injury and/or equipment damage.
Olympus has not tested the efficacy of cleaning and high-level disinfection on this equipment in
combination with endoscopes that are not listed on the “List of compatible Endoscopes/Connecting
Tubes <OER-Elite>”.
Warnings and cautions
Follow the warnings and cautions given below when handling this equipment. This information is
supplemented by warnings given in each chapter.
• Do not insert an EndoTherapy accessory or other object through an opening
including the air vent of the equipment. Also, do not allow any liquid (including
water or disinfectant solution) to flow into an opening. Contact with an electrical part
inside the equipment could cause an electric shock or equipment failure.
• Always remove the tank from the detergent/alcohol drawer before putting detergent
or alcohol in the tank. If detergent or alcohol is spilled on the detergent/alcohol
drawer, it could get inside the equipment and contact an electrical part inside,
causing an electric shock or fire hazard.
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor
or moderate injury. It may also be used to alert against unsafe practices or potential
equipment damage.
Indicates additional helpful information.
WARNING
CAUTION
NOTE