Olympus OER-Elite - Chapter 2 Nomenclature and Functions; 2.1 Front panel

Olympus OER-Elite
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.6 Reprocessing basin
27
OER-Elite INSTALLATION MANUAL
Ch.2
No. Nomenclature Description
1 Connector A2 (Light blue) Connect a connecting tube colored light blue.
This connector is mainly used for second endoscope.
2 Connector B2 (Blue) Connect a connecting tube colored blue.
This connector is mainly used for second endoscope.
3 Connector B1 (Blue) Connect a connecting tube colored blue.
This connector is mainly used for first endoscope.
4 Connector A1 (Light blue) Connect a connecting tube colored light blue.
This connector is mainly used for first endoscope.
5 Connector D2 (Orange) Connect a connecting tube colored orange.
This connector is mainly used for second endoscope.
6 Connector C2 (Green) Connect a connecting tube colored green.
This connector is mainly used for second endoscope.
7 Connector C1 (Green) Connect a connecting tube colored green.
This connector is mainly used for first endoscope.
8 Connector D1 (Orange) Connect a connecting tube colored orange.
This connector is mainly used for first endoscope.
9 Circulation port Aspirates the cleaning fluid or disinfectant solution for circulation during cleaning
or disinfection.
10 Drain port Drains liquid from the reprocessing basin.
11 Leak test connector E2
(Purple)
Connect the leak test air tube that is connected to the second endoscope here.
12 Leak test connector E1
(Purple)
Connect the leak test air tube that is connected to the first endoscope here.
13 Washing case mount Mount for the Washing case that is used to hold the endoscope accessories
including valves for reprocessing.
14 Temperature sensor Monitor the temperature of the fluid in the reprocessing basin.
15 Water supply piping
disinfection connector
Connect the water supply piping disinfection hose here.
16 Fluid level sensor Detects abnormal fluid level in the reprocessing basin.
17 Disinfectant solution
nozzle
Supplies disinfectant solution to the reprocessing basin. Also, used to check if
the disinfectant concentration meets the recommended level by the disinfectant
manufacturer.(Open the cap and insert the chemical indicator (test strip).)
18 Float switch (long) Detects the fluid level in the reprocessing basin to control it.
19 Detergent nozzle Supplies detergent to the reprocessing basin.
20 Float switch (short) Detects the fluid level in the reprocessing basin to control it.
21 Circulation nozzle Supplies the cleaning fluid or disinfectant solution aspirated through the
circulation port for circulation.
22 Water supply/circulation
nozzle
Supplies water for cleaning. Also, supplies the cleaning fluid or disinfectant
solution aspirated through the circulation port for circulation.
23 Water level scale (× 3) Marks for confirming that the equipment is level.