5-46
5-10 User Parameters
5
Operating Functions
 This parameter is enabled when the Gain Switch Setting (Pn31) is set to 3, or 5 to 10. Set the delay
time from the moment the condition set in the Gain Switch Setting (Pn31) is not met until returning
to gain 1.
 This parameter is automatically changed by executing realtime autotuning function. To set it
manually, set the Realtime Autotuning Mode Selection (Pn21) to 0.
 This parameter is enabled when the Gain Switch Setting (Pn31) is set to 3, 5, 6, 9, or 10. Set the
judgment level for switching between gain 1 and gain 2.
 The unit for the setting depends on the condition set in the Gain Switch Setting (Pn31).
 This parameter is automatically changed by executing realtime autotuning function. To set it
manually, set the Realtime Autotuning Mode Selection (Pn21) to 0.
 Set the hysteresis width above and below the judgment level set in the Gain Switch Level Setting
(Pn33).
 The Gain Switch Hysteresis Setting (Pn34) is defined as shown in the following figure.
 This parameter is automatically changed by executing realtime autotuning function. To set it
manually, set the Realtime Autotuning Mode Selection (Pn21) to 0.
Pn32
Gain Switch Time
Setting range 0 to 10000 Unit × 166 µs Default setting 30
Power OFF → ON
---
Position
Pn33
Gain Switch Level Setting
Setting range 0 to 20000 Unit --- Default setting 600
Power OFF → ON
---
Pn34
Gain Switch Hysteresis Setting
Setting range 0 to 20000 Unit --- Default setting 50
Power OFF → ON
---
Position
Position
Pn33
Pn34
0
Pn32
Gain 1
Gain 1
Gain 2