3.1.2. Ambient Temperature Compensation
In dependence on the emissivity value of the object a certain amount of ambient radiation will be reflected from
the object surface. To compensate this impact, the software provides the feature Ambient control:
Internal (Head):
The ambient temperature will be taken from the head-internal Pt1000 probe
(factory default setting).
Fixed value:
A fixed value can be entered in the edit box Fixed value (if the ambient
radiation is constant).
Note
Especially if there is a big difference between the ambient temperature at the process and head
temperature the use of Ambient control with Fixed value or via the I/O pins (mode: External Ambient
Compensation).is recommended.