-86 CSvision -
4.2.2. Ratio Mode - Smart Ratio
While the standard mode requires a
constant emissivity ratio/slope, the
Smart Ratio measurement allows a data
set of different slopes to be recorded and
applied for temperature calculation. This
is required, for example, if the degree of
contamination of the protective window
changes during the process and the ratio
temperature is no longer correct. This
cannot be described with a constant
slope.
A requirement for the measurement is that the object temperature must be known.
Note: Before the first use a data record must be recorded
In the Teach-in function, the data records are recorded.
The Smart Ratio mode can be activated or deactivated with the On/Off buttons.