CSvision 97-
Averaging
In this mode an arithmetic algorithm will be performed to smoothen the
signal. The Averaging Time is the time constant. This function can be
combined with all other post processing functions. The minimum adjustable
average time is 0,001 s.
Peak hold
In this mode the sensor is waiting for descending signals. If the signal
descends the algorithm maintains the previous signal peak for the specified
Hold time.
The minimum adjustable hold time is 0,001 s.
After the hold time the signal will drop down to the second highest value or
will descend by 1/8 of the difference between the previous peak and the
minimum value during the hold time. This value will be held again for the
specified time. After this the signal will drop down with slow time constant
and will follow the current process temperature.
Therefore, if periodic events will be measured (bottles on a conveyor e.g.)
this peak hold function avoids a drop down of the signal to the conveyor
temperature in-between 2 events.
Valley hold
In this mode the sensor waits for ascending signals. If the signal ascends
the algorithm maintains the previous signal valley for the specified Hold
time. The definition of the algorithm is according to the peak hold algorithm
(inverted).
Advanced Peak hold
In this mode the sensor waits for local peak values. Peak values which are
lower than their predecessors will only be taken over if the temperature has
fallen below the Threshold value beforehand. If Hysteresis is activated a

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the optris CompactPlus Connect and is the answer not in the manual?

optris CompactPlus Connect Specifications

General IconGeneral
Brandoptris
ModelCompactPlus Connect
CategoryThermometer
LanguageEnglish

Related product manuals