DIS-26
SECTION 2 DISASSEMBLY PROCEDURES
<RZ31K/RS30K>
C
A
B
B
A
D
G
E
H
F
H
J
2. 23. DMD block/liquid cooling unit removal
1. Loosen 6 screws-A and remove the DMD top cover.
2. Loosen 4 screws-B and remove the prism fan block.
3. Unscrew 1 screw-C, 4 screws-D and remove the color prism 2 fan and 3 fan.
Prism fan block
DMD cover block
Color prism 2 fan
Color prism
3 fan
4. Unscrew 1 screw-E, 1 screw-F and remove DMD support metal.
5. Unscrew 4 screws-G, 4 screws-H, 4 screws-J and remove the DMD block/intake duct block upward.
Color prism
4 fan
[RE]-P.C.Board
[M2]-P.C.Board
Color prism 1
fan
DMD block/intake duct block
A
B
C
D
E
J
F
G
H
DMD support metal
Note: When removing the DMD block/intake duct block, be
careful not to hook up the lead wires.