45
Then a window will open where you need to enter the number of hardness measures
for calibration (Fig. 2.37), use the buttons
to enter the desired number of measures, and press .
Figure 2.37 - Enter the number of measures to calibrate
A window will appear (Fig. 2.38), where measurements should be made on calibration
measures one after another. After making several impacts on one of the measures, press the
button and in the window (Fig. 2.39) use the buttons
to enter the specific value of the measure and press . Then you need
to repeat the same steps on the remaining hardness measures.
Figure 2.38 - Calibration of the device on calibration measures