35
13. [CONFIRM] CCD BASE PLATE SUPPORT PILLAR
[Required equipment] Block gauge for 35mm, Dial gauge comparator, etc. (same as *istD or K100D )
① Measure height of the CCD base support pillar (3 places) from the mount surface as shown in the figure.
Tolerance (For 76830 only): 49.45±0.1 mm
With using Block gauge for 35mm (45.46mm):
+3.89~4.09 mm
14. 0-C000 (SR/CCD block)
[Caution-1] CCD/SR block has strong magnet inside, therefore caution for adsorption of parts.
[Caution-2]Since a performance is spoiled, a CCD/SR block cannot be disassembled and also don't apply
the external pressure to a movable part
[Caution-3] The flex from SR/CCD should be take care carefully, otherwise it will effect the performance of
SR function.
[Caution-4] There is no scratch or dust on the CCD surface.
*To prevent the damage of SI-LED, attach mount cover. Then put downward.
① A32 Coil spring x3 – install to pillar.
② [Confirmation] Center plate must move up and down and right and left by own weight.
③ Arrange lead wire and flex then install CCD/SR block according to the guide pin of the plate.
[Caution] A32 must not come off and adsorption to magnet.
④ A31 adjusting screw x3 – Tighten A31 until stop while pressing to main body.
⑤ [Adjustment] Height of 0-C000, temporally adjustment by A31 x3 - Screw back two turns.
+3.
+0.89~1.09
④
③
①
②
⑤