LIBRARIES
PerkinElmer Life and Analytical Sciences 113
Quench Standards Library
The Quench Standards window allows you to enter information into and retrieve
information from the Quench Standards Library.
The data from these quench sets is used to construct quench curves for
determining DPM (Disintegrations Per Minute) in DPM Assays. Use the Quench
Standards Library to define and save the name, maximum energy value and the
number of standards for each quench set for Quench Standards Assays. Quench
standards are counted once and the entire spectrum for each quench standard is
stored in the Quench Standards Library. This allows you to select and use the
same quench set in any number of assays.
In Replay, the Quench Standards window allows you to select a quench standards
set for the purpose of reanalyzing sample data. The data from these quench sets
is used to construct quench curves for determining sample DPM (Disintegrations
Per Minute).
Note: The fields that are enabled in the Quench Standards Library
will be different when accessed from different locations within the
software.
To count a new quench curve, the following needs to be entered in the quench
standards library:
Add a name
Max keV
DPM value
# of Standards, Count Mode, Coincidence Time, Delay Before Burst, Date
Counted, and Time Counted will be filled after quench standards is counted.