MAINTENANCE AND TROUBLESHOOTING
PerkinElmer Life and Analytical Sciences 169
Warnings and Messages
System Messages
The following system messages are presented in alpha-numerical order.
2000CA File Error
The system encountered an error while writing the 2000CA.DAT file. This file is
needed by PerkinElmer application programs. Rerun the assay. If the error recurs,
call you PerkinElmer Technical Service.
? (message column)
If this character appears in the Messages column of a report, a math error
occurred when calculating the result for the sample.
< (message column)
If this character appears in the Messages column of a report, the sample was
rejected due to low count rate within the first 30 seconds of the sample count.
# (message column)
If this character appears in the Messages column of a report, the sample was
determined to be heterogeneous. Typically, this is due to phase separation in the
sample vial.
* (message column)
If this character appears in the Messages column of a report, the sample counts
exceeded 100,000 in either High Sensitivity or Low Level count mode. The
instrument automatically switched to Normal count mode.
A (message column)
If this character appears in the Messages column of a report, this data line
represents an Average value for replicate sample vials.
A quench set associated with this assays nuclide was modified after
being run.
Information for a quench set has been changed in the Quench Standards Library,
but the standard set has not been recounted after the changes were made.
Recount the Quench set.