CHAPTER 9
170 PerkinElmer Life and Analytical Sciences
A quench set associated with this assay’s nuclide was not counted
with a sufficient number of samples.
A sample nuclide has been defined in the library, but the quench set selected for
the nuclide has not been counted. A minimum of two quench standards must be
counted for each quench set.
Count the appropriate quench standards.
An active protocol is invalid for group priostat.
An active protocol was selected as a Group priostat protocol.
Select a different protocol or wait until the active protocol is completed.
B (message column)
If this character appears in the Message column of a report, this data line
represents data about a Background Subtraction vial.
BACK PIN JAM FWD
A Sample changer error has occurred.
No action is required unless a message is printed. If this message appears on a
printout, call PerkinElmer Technical Service.
BACK PIN JAM REV
A Sample changer error has occurred. No action is required unless a message is
printed.
If this message appears on a printout, call PerkinElmer Technical Service.
BAD ELEV/EXT STD
A Sample changer error has occurred. No action is required unless a message is
printed.
If this message appears on a printout, call PerkinElmer Technical Service.
Bad shutter.
A Sample changer error has occurred. No action is required unless a message is
printed.
If this message appears on a printout, call PerkinElmer Technical Service.