ADVANCED FEATURES
PerkinElmer Life and Analytical Sciences 151
Low Level Preview
The Low Level Preview function is a Sample Priostat option available via the Run
menu. This feature allows you to observe the sample spectrum and approximate
the count rate of a sample using Low Level count mode.
Figure 8-9 Sample Priostat Window
To preview sample counting using the Low Level Preview feature, you will need to
perform the following tasks:
1. Load your sample(s) into a cassette. Attach the Priostat flag in the reset
position to the cassette.
2. Select Low Level Preview from the Run menu. The Sample Priostat window is
displayed.
3. Select Count from the Run menu. Sample counting will commence immediately
after the instrument identifies the Priostat flag.
4. Stop the counting of a sample by selecting Stop from the Run menu.
5. Select Next Sample followed by Count from the Run menu if you would like to
count additional samples.
6. Stop the sample counting procedure by selecting End Sample Priostat from the
Run menu.