MAINTENANCE AND TROUBLESHOOTING
PerkinElmer Life and Analytical Sciences 179
Vial release close.
A Sample changer error has occurred.
No action is required unless a message is printed. If this message appears on a
printout, call PerkinElmer Technical Service.
Vial release open
A Sample changer error has occurred.
No action is required unless a message is printed. If this message appears on a
printout, call PerkinElmer Technical Service.
W (message column)
If this character appears in the Message column of a report, this indicates that
the Positive ID (PID) that was entered in the worklist does not match the PID read
from the cassette.
Verify that the PID number that you used in the worklist matches that found on
the appropriate cassette.
Warning: C-14 Normalization DPM not defined.
The C-14 Normalization DPM has not been defined.
In the IPA Definition window, enter a DPM value for the C-14 Standard.
Warning: H-3 Normalization DPM not defined.
The H-3 Normalization DPM has not been defined.
In the IPA Definition window, enter a DPM value for the H-3 Standard.
Warning: IPA background data not available.
No IPA background values are available for background subtraction.
Perform the IPA procedures.
Warning: questionable C-14 background value-please view historic
data.
The background source failed the IPA check in the Carbon-14 region.
If Carbon-14 background is high and Tritium background is normal, there may be
an increase in environmental radiation. The detector may be contaminated, or
light is leaking into the detector. Check the background spectrum via the
SpectraView window. Review the historic IPA data. Call PerkinElmer Technical
Service.