MAINTENANCE AND TROUBLESHOOTING
PerkinElmer Life and Analytical Sciences 181
Warning: questionable H-3 figure of merit value-please view historic
data.
The Figure-of-Merit value for Tritium is below the threshold entered in the IPA
Definitions window.
Check the Figure-of-Merit threshold value. Check the efficiency and background
values for Tritium. If these values are within specifications, call PerkinElmer
Technical Service.
Warning: system not normalized.
The instrument is not normalized.
Perform the instrument Calibration and Normalization procedures.
Warning: User Has Modified Count Conditions
Changes have been made to the assay that is currently counting.
Allow the current assay to finish counting prior to making changes.
You cannot exit the application because the instrument is counting a
protocol.
The QuantaSmart program cannot be closed while the instrument is counting
samples.
Allow the instrument to finish counting prior to closing the program.
You cannot exit the application because the instrument is counting
group priostat.
The QuantaSmart program cannot be closed while the instrument is counting
samples.
Allow the instrument to finish counting prior to closing the program.
You cannot exit the application because the instrument is engaged in
a sample priostat activity.
The QuantaSmart program cannot be closed while the instrument is counting
samples.
Allow the instrument to finish counting prior to closing the program.