2. Click one of the Quench Set buttons. The Quench Standards window is
displayed.
Figure 2-19 Quench Standards Window.
3. Select the Quench Set Low if you are counting one nuclide in one counting
region, the Quench Set Medium if you are counting a second radionuclide in a
separate region and the Quench Set High if you are counting a third
radionuclide in a third region.
4. Select the name of the quench set you would like to link to the sample nuclide.
5. Click OK. The name of the quench set(s) you selected should appear in the
Sample Nuclides Library window on the Quench Set buttons.