THE SYSTEM SOFTWARE
PerkinElmer Life and Analytical Sciences 53
Quench Standards
The Quench Standards window allows you to define a new quench set to be
stored or select a quench standards set for DPM assays or data reanalysis in
Replay. The data from these quench sets is used to construct quench curves for
determining sample DPM (Disintegrations Per Minute).
Note: Name, max keV and DPM are the only entries required when
entering a new quench set. The remaining parameters stored in
the library are retrieved from the assay parameters chose in
defining an assay. The number of standards (up to twenty) is
determined automatically during counting.
For additional information, refer to the Libraries chapter.
Figure 4-24 Quench Standards Window