ASSAYS
PerkinElmer Life and Analytical Sciences 73
Performing an Alpha Beta Assay
The following tasks are required when performing an Alpha Beta Assay.
Calibrate the Instrument, if necessary.
Define and Run an Alpha Beta Standards Assay (page 75) to establish the
optimal pulse decay discriminator value.
Define an alpha beta nuclide in the Alpha Beta Nuclide Library. Choose a
Standards Set (second column in the table) from the Alpha Beta Standards
Library to use the discriminator setting from the Standard Set. If you do not
choose a Standard Set, manually enter a discriminator setting directly in the
Alpha Beta Nuclide Library.
Open an existing assay or create a new assay by choosing Alpha Beta assay
type. Select the desired alpha beta radionuclide name on the Count
Conditions tab and define the remaining assay parameters
Save the assay in the Assays folder of the Packard\TriCarb directory.
Associate (link) the assay parameters with a protocol number in the protocol
tree and attach the corresponding protocol clip to the first cassette to be
counted.
Load the cassette(s) with vials and load the instrument with cassettes.
Start the instrument.