90 . Spectrum 100 Series User’s Guide
3. Remove the beamsplitter retaining screw.
4. Noting its installed position, pull out the beamsplitter.
5. Install the new beamsplitter and secure with the retaining screw.
6. Refit the laser baffle and secure with retaining screws.
7. Close and secure the Main Cover (steps 10 to 15 on page 86).
8. Plug in and switch on the instrument.