Message Conventions
LAN/WLAN Data Export
2-8 Efficia CM Series Data Export Guide
3.2 universal ID The identifier
for all Efficia
monitors.
ST 20 R 1 1 EfficiaCM
7 Date/Time
of Message
The date and
time that the
monitor
created the
message.
TS 26 R 1 1
7.1 Date/Time Timestamp in
the format
YYYYMMDDHH
MMSS.
NM 20 R 1 1 20140510124500
9 Message
Type
The server uses
this field to
know which
data segments
to recognize.
Contains the
message type,
trigger event,
and message
structure ID.
CM_MSG 15 R 1 1 0076 ORU^R01^ORU_R01
9.1 message
type
Unsolicited
Observation
Reporting
(ORU)
message.
ID 3 R 1 1 0076 ORU
9.2 trigger
event
Unsolicited
transmission of
results: R01.
ID 3 R 1 1 0003 R01
9.3 message
structure
ID 7 R 1 1 0354 ORU_R01
10 Message
Control ID
The unique
identifier for
this message
instance. The
identifier is a
combination of
the
10-character
device serial
number and a
10-character
message
counter.
ST 20 R 1 1 MONITORABC000000002
D
11 Processing
ID
Indicates
whether to
process the
message as
part of a
production (P),
training (T), or
debugging (D)
system.
PT 3 R 1 1
Seq Field Description DT Len Usage Min Max Table Sample Value