Serial Data Export
Efficia CM Series Data Export Guide 3-13
Observation Reporting (ORF) Message Syntax
The following table describes each field in the MSA segment.
Query Definition (QRD) Segment
The QRD segment in the ORF message is the same QRD segment that was sent in the
QRY message.
QRD|20140510124500|R|I|QUERY123|||||RES<CR>
This QRD segment shows the query message being sent to the monitor on May 10, 2014,
at 12:45:00 pm. The receiver requests the status of all parameters in the numeric panes
on the monitor.
QRD|20140510124500|R|I|QUERY123|||||STA<CR>
The following table describes each field in the QRD segment.
Seq Field Description DT Len Usage Min Max Table Sample Value
1 Acknowledg
ement Code
Contains the
acknowledg-
ment code. AA =
Application
Accept (positive
acknowledg-
ment).
ID 2 R 1 1 AA
2 Message
Control ID
Contains the
message control
ID of the QRY
message sent by
the server.
ST 20 R 1 1 EXAMPLE123
Seq Field Description DT Len Usage Min Max Table Sample Value
1 Query Date/
Time
Contains the
date and time
from the
incoming QRY
message; this is
used for the
ORF response
message.
TS 26 R 1 1 20140510124500
2 Query
format code
The format of
the response.
Must always be
record-oriented
(R).
ID 1R 11 R
3 Query
Priority
The timeframe
in which the
response is
expected. Must
always be
Immediate (I).
ID 1R 11 I
4 Query ID A unique
identifier for the
query.
ST 10 R 1 1 QUERY123