Serial Data Export
Efficia CM Series Data Export Guide 3-7
Query (QRY) Message Syntax
Query Definition (QRD) Segment
The QRD segment specifies the date and time of the query, the format code, and the
priority. The QRD segment is sent with the QRY message and then returned in the ORF
message.
Examples:
This QRD segment shows the query message being sent to the Efficia CM Series monitor
on May 10, 2014, at 12:45:00 pm. The receiver requests the status of all parameters in the
numeric panes on the monitor.
QRD|20140510124500|R|I|QUERY123|||||STA<CR>
11 Processing
ID
Indicates
whether to
process the
message as
part of a
production
(P), training
(T), or
debugging (D)
system.
PT 3 R 1 1
11.1 processing
ID
All results
from the
monitor are
production (P)
messages.
ID 3 R 1 1 0103 P
12 Version ID The HL7
version that
the server
matches to its
own version to
be sure the
message will
be interpreted
correctly. HL7
is backwards
compatible;
any previous
version is
compatible
with this
specification.
VID 8 R 1 1 0104
12.1 version ID Version 2.4 is
currently
supported.
ID 3 R 1 1 0104 2.4
Seq Field Description DT Len Usage Min Max Table Sample Value