Positive ACK Message Syntax
Serial Data Export
3-26 Efficia CM Series Data Export Guide
Message Acknowledgement (MSA) Segment
The MSA message contains the response code and the initiating message identifier.
Example:
This MSA segment shows that the server successfully received and validated the ORF
message from the monitor.
MSA|AA|MONITORABC000000002D<CR>
The following table describes each field in the MSA segment.
11 Processing
ID
Indicates
whether to
process the
message as part
of a production
(P), training (T),
or debugging (D)
system.
PT 3 R 1 1
11.1 processing
ID
All results from
the monitor are
Production (P)
messages.
ID 3 R 1 1 0103 P
12 Version ID The HL7 version
that the server
matches to its
own version to
be sure the
message will be
interpreted
correctly. HL7 is
backwards
compatible; any
previous version
is compatible
with this
specification.
VID 8 R 1 1 0104
12.1 version ID Version 2.4 is
currently
supported.
ID 3 R 1 1 0104 2.4
Seq Field Description DT Len Usage Min Max Table Sample Value
Seq Field Description DT Len Usage Min Max Table Sample Value
1 Acknowledg
ement Code
Contains the
acknowledg-
ment code. The
following codes
are supported:
AA = Accept
AR = Reject
AE = Error
ID 2 R 1 1 0008 AA
2 Message
Control ID
Contains the
message control
ID of the
initiating
message (ORF).
ST 20 R 1 1 MONITORABC000000002
D

Table of Contents