MDIL Coding
LAN/WLAN Data Export
2-26 Efficia CM Series Data Export Guide
The following table lists MDIL codes for Pulse Rate source.:
Alarms
The following MDIL-ALERT codes identify the severity of an alarm on the Efficia CM
Series monitor:
0004-0c00 l/min Liters per minute
0004-06c3 kg Kilogram
0004-06e0 lb Pound
0004-0511 cm Centimeter
0004-0560 in Inch
MDIL Code Parameter Label Description
MDIL Code Description
0002-4a07 NBPm Mean NBP Pulse Rate source
0002-4bb8 SpO2 SpO
2
Pulse Rate source
MDIL-ALERT Code Description
6 Soft Inop Low priority (technical) alarm on the monitor.
2 Yellow Alarm Medium priority alarm on the monitor.
1 Red Alarm High priority alarm on the monitor.