LAN/WLAN Data Export
Efficia CM Series Data Export Guide 2-11
Message Conventions
Common Order (ORC) Segment
The ORC segment in the ORU message transmits the monitor name.
Example:
This ORC segment contains the monitor name MONITORABC.
ORC|NW|||||||||||||||||MONITORABC<CR>
The following table describes each field in the ORC segment.
Observation Request (OBR) Segment
The OBR segment in the ORU message identifies the type of data being exported, the
date and time that the measurements occurred, and the ID of the person taking the
measurements. If the monitor is sending trend data, the data type in the OBR segment is
MONITOR. If the monitor is sending an alarm, the data type in the OBR segment is
ALARM.
Examples:
This OBR segment identifies the type of data as MONITOR. The measurements occurred
at 12:45:00 pm on May 10, 2014 and were collected by OPERATOR12.
OBR||||MONITOR|||20140510124500|||OPERATOR12<CR>
19 Visit Number Unique number
assigned to the
patient visit.
CX 250 RE 0 1
19.1 ID Transaction or
visit ID.
ST 20 RE 0 1 TRANSACTION1234
39 Department Department to
which patient is
admitted
ST 20 RE 0 1
44 Admit Time Time when
patient is
admitted in the
format
YYYYMMDDHH
MMSS
NM 14 RE 0 1
Seq Field Description DT Len Usage Min Max Table Sample Value
Seq Field Description DT Len Usage Min Max Table Sample Value
1 Order
Control
The function of
the order
segment. For
example,
NW = new order.
ID 2 R 1 1 0119 NW
18 Entering
Device
Identifies the
physical device
(monitor) used
to enter the vital
signs
measurements.
CE 250 R 1 1
18.1 ID Monitor name. ST 10 R 1 1 MONITORABC