Serial Data Export
Efficia CM Series Data Export Guide 3-17
Observation Reporting (ORF) Message Syntax
Observation/Result (OBX) Segment
Each OBX segment in the ORF message reports a single measurement. There can be
multiple OBX segments in an ORF message.
Examples:
These OBX segments contain the following measurements: SpO
2
(98%); systolic NBP (122
mmHg); diastolic NBP (85 mmHg); mean NBP (98 mmHg); Pulse Rate (80 bpm) derived
from SpO
2
; ECG respiration (20 rpm); Predictive temperature (37.4°C). All of the results
are final.
OBX||NM|0002-4bb8^SpO2^MDIL||98|0004-0220^%^MDIL|||||F<CR>
OBX||NM|0002-4a05^NBPs^MDIL||122|0004-
0f20^mmHg^MDIL|||||F|||20120510123500<CR>
OBX||NM|0002-4a06^NBPd^MDIL||85|0004-
0f20^mmHg^MDIL|||||F|||20120510123500<CR>
OBX||NM|0002-4a07^NBPm^MDIL||98|0004-
0f20^mmHg^MDIL|||||F|||20120510123500<CR>
OBX||NM|0002-480a^Pulse^MDIL||80|0004-0aa0^bpm^MDIL|||||F|||||
|0002-4bb8^SpO2^MDIL<CR>
OBX||NM|0002-5000^Resp^MDIL||20|0004-0ae0^rpm^MDIL|||||F<CR>
OBX||NM|0002*f0c7^Temp^MDIL||37.4|0004-1710^C^MDIL|||||F||||||0002-
f110^Oral^MDIL<CR>
In these OBX segments, the empty Observation Value field (OBX-5) indicates that there is
no NBP measurement on the monitor screen:
OBX||NM|0002-4a05^NBPs^MDIL|||0004-0f20^mmHg^MDIL|||||F<CR>
OBX||NM|0002-4a06^NBPd^MDIL|||0004-0f20^mmHg^MDIL|||||F<CR>
OBX||NM|0002-4a07^NBPm^MDIL|||0004-0f20^mmHg^MDIL|||||F<CR>
In this OBX segment, the X in the Observation Result Status field (OBX-11) indicates an
invalid SpO
2
measurement, which is displayed as a question mark (?) on the monitor
screen.
4.1 identifier Identifies the
type of result.
Valid values are:
MONITOR
ST 20 R 1 1 MONITOR
7 Observation
Date/Time
Date and time of
the
measurement.
TS 26 R 1 1
7.1 Date/Time Timestamp in
the format
YYYYMMDDHH
MMSS.
NM 20 R 1 1 20140510124500
10 Collector
Identifier
Identifies the
person collecting
the
measurements.
XCN 20 RE 0 1
10.1 ID number The ID of the
person collecting
the
measurements.
ST 12 RE 0 1 OPERATOR12
Seq Field Description DT Len Usage Min Max Table Sample Value