Loading...
Observation Reporting (ORF) Message Syntax
Serial Data Export
3-10 Efficia CM Series Data Export Guide
Note — Each message segment consists of a variable number of lines, ending with a
carriage return character. In the examples, long lines are broken for display
purposes, and the carriage return character is indicated by <CR>.
Examples:
The following example shows a ORF message sent by an Efficia CM Series monitor to
report the status of all parameters in the numeric panes.
MSH|^~\&|^EfficiaCM||||20140510124500||ORF^R04|MONITORABC000000002D|P
|2.4|||||||||2.16.840.1.113883.9.2.1<CR>
MSA|AA|EXAMPLE123<CR>
QRD|20140510124500|R|I|QUERY123|||||STA<CR>
PID|||MRN1234||JONES^MARY^A<CR>
PV1|||^^LOCATION34|||||||||||||||A|TRANSACTION1234<CR>
ORC|NW|||||||||||||||||MONITORABC<CR>
OBR||||MONITOR|||20140510124500|||OPERATOR12<CR>
OBX||NM|0002-4bb8^SpO2^MDIL||98|0004-0220^%^MDIL|||||F<CR>
OBX||NM|0002-4a05^NBPs^MDIL||122|0004-
0f20^mmHg^MDIL|||||F|||20140510124430<CR>
OBX||NM|0002-4a06^NBPd^MDIL||85|0004-
0f20^mmHg^MDIL|||||F|||20140510124430<CR>
OBX||NM|0002-4a07^NBPm^MDIL||98|0004-0f20^mmHg^MDIL|||||F<CR>
OBX||NM|0002-480a^Pulse^MDIL||80|0004-0aa0^bpm^MDIL|||||F|||||
|0002-4bb8^SpO2^MDIL<CR>
OBX||NM|0002-5000^Resp^MDIL||20|0004-0ae0^rpm^MDIL|||||F<CR>
OBX||NM|0002-f0c7^T1^MDIL||37.4|0004-1710^C^MDIL|||||F<CR>
Message Header (MSH) Segment
Every ORF message begins with the MSH segment, which contains message envelope
information, such as the source, destination, and some specifics of the message syntax.
The MSH segment does not contain clinical information.
Example:
This MSH segment shows an ORF message being sent from the Efficia CM Series series
monitor on May 10, 2014, at 12:45:00 pm.
MSH|^~\&|^EfficiaCM||||20140510124500||ORF^R04|MONITORABC000000002D|P|
2.4|||||||||2.16.840.1.113883.9.2.1<CR>
OBX Observation/Result R 1 no max
Segment Description Usage Min Max