Query (QRY) Message Syntax
Serial Data Export
3-8 Efficia CM Series Data Export Guide
The following table describes each field in the QRD segment.
Seq Field Description DT Len Usage Min Max Table Sample Value
1 Query Date/
Time
The query date
and time. This
field is ignored
by the monitor,
but the given
value is sent to
the server in the
ORF message.
Note — The
server should
supply this value
in order to be
HL7 compliant.
TS 26 R 1 1 20140510124500
2 Query
format code
Must always be
R for record-
oriented, The
monitor does not
check the format
code.
ID 1 R 1 1 R
3 Query
Priority
Must always be
Immediate.(I).
ID 1 R 1 1 I
4 Query ID A unique
identifier. This
field must not be
empty. The
Query ID is not
processed, but is
sent to the serve
in the ORF
message.
Note — The
monitor does not
check whether
query IDs are
unique.
ST 10 R 1 1 QUERY123
7 Quantity
Limited
Request
Always empty for
NO LIMIT. This
field is ignored.
CQ 10 0 0
9 What
subject filter
Tilter must be the
same in query
and response.
Valid values are:
STA =
Status,
screen
display
CE 60 R 1 1
STA