Message Conventions
LAN/WLAN Data Export
2-16 Efficia CM Series Data Export Guide
Alarms
This section describes the OBX segment that reports alarms.
Example:
This OBX segment contains a medium-level physiological alarm.
OBX||TX|2^Yellow Alarm^MDIL-ALERT||HR Low, HR=55 /min||||||F<CR>
The following table describes the fields in the OBX segment that reports alarms.
17.3 name of
coding
system
For Heart Rate
source, Pulse
Rate source, and
temperature
method, the
default coding
system is MDIL.
For more
information, see
“MDIL Coding”
on page 2-20.
IS 20 RE 0 1 0396 MDIL
Seq Field Description DT Len Usage Min Max Table Sample Value
Seq Field Description DT Len Usage Min Max Table Sample Value
2 Value Type Format of the
alarm in the
Observation
Value field.
The format is TX
(text).
ID 2 R 1 1 0125 TX
3 Observation
Identifier
Contains a
unique identifier
for the alarm.
The format is:
<identifier>^
<text>^
<name of
coding
system>
CE 250 R 1 1 2^Yellow
Alarm^MDIL-ALERT
3.1 identifier Code that
uniquely
identifies the
alarm.
For alarms, the
identifier is an
MDIL-ALERT
code. For more
information, see
“MDIL Coding”
on page 2-20.
ST 20 R 1 1 2
3.2 text Description of
the alarm
priority.
ST 20 R 1 1 Yellow Alarm