Error Codes
Troubleshooting
3-32 Efficia CM Series Repair Guide
1029
System Error (1029)
Could not allocate
enough memory.
Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See If the problem
persists, replace the front end
assembly. See “Removing the Front
End Assembly” on page 4-20.
1030
System Error (1030)
An error occurred
when the system shut
down front end
devices.
Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See “Replacing the
Main Board” on page 4-27.
If the problem persists, replace the
front end assembly. See “Removing
the Front End Assembly” on
page 4-20.
1031
System Error (1031)
The monitoring and
watchdog thread
could not initialize.
Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See If the problem
persists, replace the front end
assembly. See “Removing the Front
End Assembly” on page 4-20.
1032
System Error (1032)
Could not open audio
files.
Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See “Replacing the
Main Board” on page 4-27.
1033
System Error (1033)
Unspecified error. Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See “Replacing the
Main Board” on page 4-27.
1035
System Error (1035)
Cannot initialize the
front-end manager.
Shut down (see “Shutting Down the
Monitor” on page 4-1) and restart
the monitor.
If the problem persists, replace the
main board. See “Replacing the
Main Board” on page 4-27.
Code Description Possible Cause Action