EasyManua.ls Logo

Philips V60 - Page 123

Philips V60
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms, messages, and troubleshooting
9-9
Low Minute Ventila-
tion
Estimated minute ventila-
tion is less than the Lo
E
setting. Escalates to a
high-priority alarm if the
alarm condition persists
for more than 60 sec.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Low
(56)
Yes Yes Yes
Low O
2
Supply Pres-
sure
O
2
inlet pressure
is less
than 30 psig and deliv-
ered O
2
is at least 5%
lower than O
2
setting. The
ventilator continues to de-
liver as much O
2
support
as possible, but it ends O
2
support when the O
2
inlet
pressure drops to less
than 18 psig. Autoresets
when O
2
inlet pressure ex-
ceeds 23 psig.
Check the patient. Attach
to oxygen source with suf-
ficient pressure.
If problem persists, pro-
vide alternative ventila-
tion. Have the ventilator
serviced.
High
(48)
No Yes Yes
Low Rate A low-priority alarm if the
measured respiratory rate
is less than the Lo Rate
setting, escalating to a
high-priority alarm in 60
sec.
A high-priority alarm from
the start if:
The Lo Rate setting is
4 BPM and there are no
breaths for > 60/Lo Rate
setting.
The Lo Rate setting is >
4 BPM and there are no
breaths for > 15 sec.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Low/
High
(55)
Yes Yes Yes
Low Tidal Volume Estimated tidal volume is
less than the Lo V
T
set-
ting. Escalates to a high-
priority alarm if the alarm
condition persists for
more than 60 sec.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Low
(57)
Yes Yes Yes
Mask:x, Exh Port:y
Use Menu to change
Displays when ventilator is
turned on. Displays select-
ed mask type and exhala-
tion port.
Select mask and port from
Menu tab. Message is re-
moved when user con-
firms selections, or after 5
minutes.
Infor-
mation
(64)
No Yes N/A
Table 9-3: Alarm and other messages: summary and troubleshooting (continued)
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
V

Table of Contents

Other manuals for Philips V60

Related product manuals