11-1
Chapter 11. Technical specifications
Control settings Table 11-1 lists ventilator control setting ranges, resolutions, and accuracies.
Table 11-2 lists the controls active in the different ventilation modes.
Table 11-1: Control setting ranges, resolutions, and accuracies
Parameter Range Resolution Performance Accuracy Factory default
Mode settings
Modes AVAPS, CPAP, S/T, PCV,
PPV (optional)
N/A N/A S/T
Control settings
C-Flex OFF, 1 to 3 1 N/A 2
CPAP 4 to 25 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
4 cmH
2
O
EPAP 4 to 25 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
4 cmH
2
O
Flow
(High flow therapy,
3.00 software and
above, and V60 Plus)
10 to 80 L/min 5 L/min N/A 35 L/min
IPAP 4 to 40 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
12 cmH
2
O
I-Time (Inspiratory
Time)
0.30 to 3.00 sec 0.05 sec ± 0.03 sec 1.00 sec
Max E 0 to 100 cmH
2
O/L 1 cmH
2
O/L N/A 15 cmH
2
O/L
Max P (PPV Maximum
Pressure Limit)
5 to 40 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
20 cmH
2
O
Max P (AVAPS Maxi-
mum IPAP Pressure)
6 to 40 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
25 cmH
2
O
Max R 0 to 50 cmH
2
O/L/sec 1 cmH
2
O/L/s N/A 4 cmH
2
O/L/s
Max V (PPV Maximum
Volume Limit)
200 to 3500 mL 5 mL ± 15% 1000 mL
Min P (AVAPS Mini-
mum IPAP Pressure)
5 to 30 cmH
2
O1 cmH
2
(2 cmH
2
O + 4% of
target)
10 cmH
2
O
O
2
(Oxygen) 21 to 100% 1% ± 5% 21%
PPV % 0 to 100% 1% N/A 30%