Technical specifications
11-7
Pneumatic
specifications
Table 11-9: Pneumatic specifications
Parameter Specification
High-pressure oxygen supply Connector: DISS male, DISS female, NIST
Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.31 to 6.00 bar / 331 to 600 kPa / 48 to
87 psig
Flow: 175 SLPM
High-pressure oxygen supply
(using V60/V60 Plus manifold)
Connector: DISS male, DISS female, NIST
Pressure: 3.10 to 6.00 bar / 310 to 600 kPa / 45 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.66 to 6.00 bar / 366 to 600 kPa / 53 to
87 psig
Flow: 175 SLPM
Air supply Integrated blower
Inspiratory outlet
(to patient port)
Connector: ISO 15 mm female/22 mm male conical
Maximum limited pressure
(P
LIMmax
)
64 cmH
2
O
Maximum working pressure range
(P
Wmax
)
5 to 50 cmH
2
O, ensured by High Inspiratory Pressure
(HIP) alarm limit
Subatmospheric pressures
generated during exhalation
None