E-12
Diagnostic mode
Service NOTE: All ventilator mode and alarm settings, alarm messages and
significant events are automatically logged and retained, even when
power is lost.
The Service screen lets you view the event log. Other service functions are for
use by authorized service personnel.
Significant Event Log
The Significant Event Log contains data about clinically relevant ventilator
occurrences, including alarms and setting changes. The time, date, and an
identifier for event classification are included. A maximum of 2,000 event
records are retained. The oldest events are overwritten first to allow recording
of new events.
1. From the Service screen, select the Misc tab.
2. The Miscellaneous screen opens (Figure E-11). Select Significant
Event Log.
Figure E-11: Miscellaneous screen