EasyManua.ls Logo

Philips V60 - Running Alarm Tests; Preparation; High Inspiratory Pressure

Philips V60
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting up the ventilator for use
5-9
d. VERIFY that the blue “Running on Internal Battery” message is
displayed.
6. Install a fully-charged battery and reconnect the ventilator to AC
power.
7. Repeat step 5 to verify that the ventilator switches over to battery
power.
Running alarm tests The ventilator performs a self-check during start-up and continuously during
operation. Alarm functionality is verified by this self-check. You may also want
to run alarm tests, which demonstrate the alarms’ operation. Follow these
steps to perform the tests.
WARNING: To prevent possible patient injury, always return alarm settings to
hospital-standard values after verifying ventilator operation.
Preparation
1. Set the ventilator up as for normal ventilation, complete with
breathing circuit (PN 582073 or the equivalent) and a 1-liter test lung
assembly (PN 1021671).
2. Set the mode to S/T and make the following control settings: Rate:
4 BPM, IPAP: 10 cmH
2
O, EPAP: 6 cmH
2
O, I-Time: 1 sec, Rise: 1,
Ramp: Off, O
2
: 21%.
3. Make the following alarm settings: Hi Rate: 90 BPM, Lo Rate: 1 BPM,
Hi V
T
: 2000 mL, Lo V
T:
OFF, HIP: 50 cmH
2
O, LIP: OFF, Lo
E
: OFF,
LIP T: 5 secs.
High Inspiratory Pressure
1. Lower the HIP alarm limit to 8 cmH
2
O.
2. VERIFY that the
High Inspiratory Pressure alarm is activated, the ventilator
cycles into exhalation, and pressure falls to 6 cmH
2
O (the EPAP
level).
3. Raise the HIP alarm limit to 15 cmH
2
O.
V

Table of Contents

Other manuals for Philips V60

Related product manuals