11-2
Technical specifications
Ramp Time OFF, 5 to 45 min 5 min ± 1 sec OFF
Rate (Respiratory Rate) 4 to 60 BPM 1 BPM ± 1 BPM 4 BPM
Rise (Rise Time) 1 to 5 1 N/A 3
V
T
(AVAPS Target Tidal
Volume)
200 to 2000 mL BTPS 5 mL ± 15% 500 mL
Table 11-1: Control setting ranges, resolutions, and accuracies (continued)
Parameter Range Resolution Performance Accuracy Factory default
Table 11-2: Controls active in Respironics V60/V60 Plus ventilation modes
CPAP S/T PCV AVAPS PPV
Timing
Rate Rate
*
I-Time I-Time*
Baseline pressure CPAP EPAP
Inspiratory pres-
sure
IPAP Max P Max P
Min P IPAP*
Rise Time
Rise Rise*
O
2
O
2
Volume V
T
Max V
Ramp feature Ramp Time
Mode-specific C-Flex PPV %
Max E
Max R
* Used in backup only